Recall Z-1939-2010 ยท DWA Control, Pump Speed, Cardiopulmonary Bypass

Medtronic Cardiovascular Revascularization & Surgical Therap

Hospital Grade Power Cord, UL, Part # 66319, 055476, manufactured by Electri-Cord Manufacturing Co. beginning in February 2003. The cords are a detachable unit, supplied with the instruments listed below: Medtronic Bio-Console 550 centrifugal pump console, Model #95180, Catalogue # 95180, E95180, R95180. Medtr

DateJun 30, 2010 (initiated Apr 21, 2010)
StatusTerminated
ReasonThe potential risks during the use of this power cord include electrical shock, delay in setup and therapy, interruption of therapy, device failure, and fires. Although Medtronic has received no reports of adverse events related to the affected cord, a failure may potentially lead to serious adverse health consequences, depending on the device, and therapy being interrupted. This Field Notificatio
Root causeNonconforming Material/Component
ActionConsignees were sent on 4/21/10 a Medtronic "Urgent Medical Device Recall Notice" letter dated April 19, 2010. The letter was addressed to their customers and described the problem and the product involved. They recommended to discontinue use of the AC power cords and to destroy them. Requested consignees to fill out the attached Medtronic Recall Certificate (FCA 1005), and return it to Medtronics.
Quantity619 USA, 80 OUS
DistributionWorldwide distribution: USA, Latin America, Middle East, Japan, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051303BIO-CONSOLE, MODEL 560
Medtronic Perfusion Systems, Jun 10, 2005
K941921BIO-MEDICUS BIO-CONSOLE MODELS 540, 540A,B,C,D,E, & 550
Medtronic Bio-Medicus, Inc., Jul 20, 1994
K972894AUTOTRANSFUSION APPARTUS (AUTOLOG)
Medtronic Blood Management, Apr 27, 1998