ClearanceBrief
Recall Z-1703-2010 ยท DZE Implant, Endosseous, Root-Form

3m Imtec Corporation

S1813OB 1.8X13mm Sendax MDI Mini Dental Implant O-Ball Prosthetic Head

DateMay 27, 2010 (initiated May 14, 2010)
StatusTerminated
ReasonProduct has Incorrect size Labeling. Actual length is 13mm. One label is correct, and one states 10mm.
Root causePackaging process control
ActionIMTEC notified consignees by phone on May 14, 2010 and by Certified Letter dated May 14, 2010.
Quantity189 Implants
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031106IMTEC SENDAX MDI AND MDI PLUS ENDOSSEOUS IMPLANT, MODELS OB-XX, SH-XX, MOB-22, MSH-XX
Imtec Corp., Aug 12, 2003