Recall Z-1957-2010 ยท MDM Instrument, Manual, Surgical, General Use
Stryker Spine
Stryker 7 MM AVS TL Trial Spacer; Catalog number: 48389407. Manufactured by: Stryker Spine SAS, Cestas, France; Distributed in the US by: Stryker Spine, 2 Pearl Court, Allendale, NJ 07401. Intended for use in the thoraco-lumbar spine.
| Date | Jul 8, 2010 (initiated May 13, 2010) |
| Status | Terminated |
| Reason | Stryker Spine has received reports of breakage of the distal end of the AVS TL Spacer Trial during the trialling step of the AVS TL procedure. |
| Root cause | Process control |
| Action | Urgent Product Recall letters, dated May 13, 2010 were sent via Federal Express to all direct accounts (branches and hospitals) with a separate letter to surgeons. The letter identified the affected product and also described the issue, potential hazards, and risk mitigations. The letter also asked customers to quarantine any affected product and return to Stryker Spine. A Customer Response Form was provided and customers should contact their Stryker Spine Sales Rep for additional information on replacement product. Questions should be directed to Tiffani Rogers on 201-760-8206. |
| Quantity | 183 |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.