Recall Z-1707-2010 ยท KRA Catheter, Continuous Flush

Angiodynamics, Inc.

PULSESPRAY INFUSION SYSTEM, 90 cm x 20 cm infusion length, Catalog No./Ref 12500402, STERILE, EO. AngioDynamics, 603 Queensbury Avenue, Queensbury, NY 12804. Infusion Catheter.

DateMay 27, 2010 (initiated May 17, 2010)
StatusTerminated
ReasonThis lot was packaged with the wrong occluding ball guidewire. The PulseSpray Catheter is 90 cm in length and requires a 109 cm occluding ball guidewire, however this lot was packaged with a 60 cm occluding ball guidewire.
Root causeProcess control
ActionAngiodynamics issued an "Urgent Medical Device Recall" notification (dated 5/11/10) to user accounts via Certified Mail. Consignees were asked to identify, segregate and return all affected product to the firm in addition to sending in a completed Reply Form via fax. For further information, contact Angiodynamics Customer Service at 1-800-772-6446.
Quantity20 units
DistributionWorldwide Distribution -- United States (AR, NY, PA, MI, MN, LA and WA), Spain, Austria and Denmark.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K905447E-Z-EM PULSED INFUSION SYSTEM
E-Z-Em, Inc., Apr 1, 1991