Recall Z-1940-2010 ยท GKZ Counter, Differential Cell

BD Biosciences, Systems & Reagents

BD Biosciences, BD FACS Sample Prep Assistant II (SPA II), blood sample preparer for flow cytometry analysis

DateJun 30, 2010 (initiated Feb 19, 2010)
StatusTerminated
ReasonSamples prepared on the BD SPA II, SN T0076 produced inaccurate sample results due to inaccurate sample dispensed as a result of the cap piercing probe
Root causeNonconforming Material/Component
ActionThe firm filed an MDR and implemented a field action consisting of a February 2010 notification letter and product replacement of affected probes. A CAPA for the probe design has been implemented. Effectivity is being tracked via mail and follow up phone call.
Quantity349 units
DistributionWorldwide distribution: USA, Australia, Belgium, China (Hong Kong), Colombia, Jamaica, Korea, Mexico, Singapore and Uruguay.