Recall Z-2337-2010 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

COULTER LH 500 Series System

DateSep 1, 2010 (initiated Feb 12, 2010)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter identified four (4) issues with the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System: Issue 1: There is a potential for misidentification to occur when the system is configured in languages other than English or Chinese. The characters #,@,[,\,],`,{,|,}, or ~ are substituted or omitted in other languages. The issu
Root causeSoftware design
ActionThe customer recall notification was initiated when Product Corrrective Actions letters with attached Customer Response form were sent on the week of February 15, 2010 to all customers who purchased the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System. The letters will provided the customers with an explanation of the problem identified and a work around. Customers with questions regarding the Product Corrective Action were instructed to call 800-526-7694 in the United States or Canada, or contact your local Beckman Coulter Representative.
Quantity1373 units in US and 147 units in Canada
DistributionThroughout the United States and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032000COULTER LH 500 HEMATOLOGY ANALYZER