Recall Z-1766-2010 ยท MSD Catheter, Hemodialysis, Implanted

Angiodynamics, Inc.

DURAMAX (TM) Hemodialysis Catheters, Stacked Tip 40 cm Straight Basic Kit, Catalog Number/REF 10302807, 510 (k) 080400, Sterile --- Manufactured in USA by AngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, NY 12804

DateJun 4, 2010 (initiated Jan 8, 2010)
StatusTerminated
ReasonThe product has the potential for a cross-lumen leak. One of the two leak testers that are used to perform the leak test on the DuraMax Catheters was found to be faulty. The tester was not able to detect cross-lumen leaks. Both testers were used during the manufacturing 100% leak test.
Root causeOther
ActionCustomers were sent URGENT MEDICAL DEVICE RECALL LETTERS (dated 1/07/10) via Certified Mail, Return Receipt Requested. Customers were instructed to return all recalled product. A return authorization number was assigned to each customer in the recall notification letter and a UPS account was provided for return freight. A reply form was provided in the recall notification letter. Questions were directed to the company sales representatives or to Customer Service at 800-772-6446.
Quantity208
DistributionNationwide and St. Thomas, VI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.