Recall Z-1898-2010 ยท NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen

Cepheid

Cepheid Xpert MRSA/SA Blood Culture Test for use with the GeneXpert Dx System, manufactured by Cepheid, Sunnyvale, CA. Medical device, qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus and methicillin-resistant Staphylococcus aures DNA directly from patient positive blood cultures. It is not intended to monitor treatment for MRSA/SA infections. Sub culturin

DateJul 1, 2010 (initiated Apr 5, 2010)
StatusTerminated
ReasonThe firm is receiving reports of inaccurate results, which may result in delay of care for seriously ill patients.
Root causeNonconforming Material/Component
ActionThe firm, Cepheid, issued a technical bulletin to customers on April 5, 2010 and an updated to the April 5 bulletin dated April 9, 2010 instructing them to perform a reflexive repeat test on any blood culture results initially determined as having SA. If further information on this please contact your Cepheid Sales Representative 1-888-336-2743 or www.CEPHEID.COM.
Quantity5,510 units
DistributionWorldwide distribution: USA and Europe, Taiwan, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082140XPERT MRSA/SA BLOOD CULTURE ASSAY
Cepheid, Sep 29, 2008