Recall Z-1777-2010 ยท BTA Pump, Portable, Aspiration (Manual Or Powered)

Ohio Medical Corporation

MoblVac Portable Wound Care Pump; an AC/DC powered portable aspirator that provides negative pressure for the application of wound drainage; Ohio Medical Corporation, Gurnee, IL 60031; Models 769600 (with Truseal), 759600 (refurbished), 769600 Rental, 769601 (rental), 769602 (evaluation pump with Truseal) and 769610 (sales unit with case).

DateJun 7, 2010 (initiated May 10, 2010)
StatusTerminated
ReasonOhio Medical has received a report of an explosive event with a MoblVac Portable Woundcare Pump housing where parts were separated while in an enclosed nylon transport bag during use.
Root causeOther
ActionOhio Medical sent Urgent Medical Device Recall letters dated 5/7/10 to the direct accounts on 5/10/10, with follow-up telephone calls on 5/11/10. The accounts were informed of the one report of an explosion of the pump, and were requested to take the pumps out of service and return the pumps to Ohio Medical. The accounts were requested to complete and fax back to Ohio Medical the enclosed acknowledgement form indicating the number of pumps being returned. Any questions were directed to Anita Christ at 847-855-6270. Follow-up letters dated 5/13/10 were sent to the accounts on 5/14/10, provi
Quantity375 units
DistributionNationwide and internationally to Qatar and Oman
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081185MOBLVAC W.C.