ClearanceBrief
Recall Z-2125-2010 ยท DQY Catheter, Percutaneous

Bard Peripheral Vascular Inc

BARD DORADO PTA BALLOON DILATATION CATHETERS, PRODUCT CODE # DR135420

DateAug 2, 2010 (initiated May 4, 2010)
StatusTerminated
ReasonThe affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft.
Root causeNonconforming Material/Component
ActionThe Bard Peripheral Vascular "URGENT MEDICAL DEVICE RECALL NOTIFICATION" with attached "RECALL AND EFFECTIVENESS CHECK FORM" was sent to the affected consignees beginning May 7, 2010 via FedEx. The recall notice informed the consignees of the reason for recall, the products affected and instructed them not to use or further distribute any affected product. The consignees were also instructed to call the firm's Recall Coordinator Silvia Carrillo at 1-800-321-4254 Option #2 Ext 2727 (M-F 7am to 4pm MST) or email at silvia.carrillo@crbard.com to facilitate the expedient return of the affected
Quantity56 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072283DORADO PTA BALLOON DILATATION CATHETER
Bard Peripheral Vascular, Inc., Sep 19, 2007