Recall Z-1704-2011 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Endoscopy

Stryker Stiletto Electrosurgical Probe (E-Probe); Model number: 0270-070-550, 0270-070-551, 0270-070-552, 0270-070-553, 0270-070-554, 0270-070-555; Manufactured by Stryker Endoscopy Puerto Rico.

DateMar 17, 2011 (initiated Feb 25, 2010)
StatusTerminated
ReasonThe heat shrink, PTFE E-Probe Tip were out of specification (OOS) in the longitudinal change (+5% -20%).
Root causeDevice Design
ActionAll Sales reps were notified of the recall on February 25, 2010. Urgent Device Removal advisory notices were also sent to all consignees.Device Removal letters identified the specific part numbers and lot numbers involved. Letters also list the potential risk to patients as a result of the issue. Customers are asked to inspect their inventory for part and lot numbers and quarantine any affected product. If recalled product is found, customers are to contact Stryker Customer Service at 800-624-4422 and request an RMS to send back recalled units. New units will be provided at no charge. The
Quantity31, 014 units
DistributionWorldwide Distribution -- US, Australia, Canada, Germany, Greece, Iberia, Korea, Switzerland, and Newbury.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052141STRYKER STILETTO ELECTROSURGICAL PROBE
Stryker Endoscopy, Oct 6, 2005