Recall Z-0588-2011 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase

Siemens Healthcare Diagnostics, Inc.

Dimension LIP Flex Reagent Cartridge, Catalog# DF55A

DateDec 10, 2010 (initiated Mar 10, 2010)
StatusTerminated
ReasonOpen Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.
Root causeOther
ActionRecalling firm issued "Urgent Field Safety Notice", dated March 2010, informing users of the problem and to transition to the revised Lipasae method.
Quantity21171
DistributionNationwide Distribution throughout the US.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952816LIPASE FLEX REAGENT CARTRIDGE
Dupont Medical Products, Jul 20, 1995