Recall Z-1883-2011 ยท MMI Immunoassay Method, Troponin Subunit

Beckman Coulter Inc.

Access Immunoassay Systems enhanced AccuTnl Reagent Kit, Part Number: A78803 The Access Accu Tri assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnl) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage. Cardiac Troponi

DateApr 4, 2011 (initiated Dec 10, 2009)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter confirmed that under certain circumstances, the following assays may produce elevated values at the low end when run in conjunction with the enhanced Access Accu Tnl assay (P/N A 78033): (1) Access Total BhCG (P/N 33500); (2) Access HYPERsensitive hTSH/Fast hTSH (P/N 33820); (3) Access hFSH (P/N 33520). The issue can occur only when the assays are
Root causeOther
ActionThe firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated December 16, 2009 to all customers. The letter described the product, problem and actions to be taken. Beckman Coulter instructed the customers to do the following mitigation to reduce the risk of encountering an erroneous result: Systems with multiple reagent pipettors (UniCel Dxl 800 and 600; UniCel DxC 880i, 860i, 680i, and 660i): 1. Designate enhanced Access AccuTnl to a different reagent pipettor (s) than Access Total BhCG/Dil-hCG2, hTSH/Fast hTSH and hFSH. 2. Discard any open Access Total BhCG, hTSH/Fas
Quantity22,032 kits
DistributionWorldwide distribution: USA and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010429ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEM # 33340, 33345
Beckman Coulter, Inc., Jun 5, 2001
K021814ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL #33340, 33345
Beckman Coulter, Inc., Sep 27, 2002