Recall Z-1997-2010 ยท ITX Transducer, Ultrasonic, Diagnostic

Toshiba American Medical Systems Inc

PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems.

DateJul 13, 2010 (initiated Jan 5, 2010)
StatusTerminated
ReasonToshiba America Medical Systems is initiating a field correction on the PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems due to an incorrect sweep direction setting. Images acquired using the endo-cavitary transducer PVT-681MV and cross-sectional images created from these volume images are reversed horizontally on the Aplio XG (SSA-790A) and the Xario XG (
Root causeOther
ActionToshiba America Medical Systems (TAMS) will issue notification to consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification. matter. Questions regarding this letter are directed to the firm at (800) 421-1968 or to your local Toshiba representative.
Quantity27
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072918MODEL SSA-680A XARIO XG DIAGNOSTIC ULTRASOUND SYSTEM, VERSION 1.00
Toshibamedical Systems Corporation, Oct 29, 2007
K092179APLIO XG DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SSA - 790A VERSION 4.0R001
Toshiba America Medical Systems, In.C, Jul 29, 2009