Recall Z-1754-2010 ยท CAH Filter, Bacterial, Breathing-Circuit

Continental Medical Labs,Inc

Xenon Diagnostic Circuit Kits labeled as "Xenon Diagnostic Circuit", "Xenon Circuit", and Needleless Xenon Diagnostic Circuit" Manufactured for RAF MEDICAL, INC., (Model number's RAF2306, RAF2304, RAF2380, RAF2300, RAF2370, RAF2372, RAF2350, RAF2341, RAF2302, RAF2340 RAF2306, RAF2304, RAF2361), or Medico-Mart, Inc. ("Xenon Curcuit" Model number's 2304, 2306). Xenon Diagnostic Circuit Kit: Used

DateJun 3, 2010 (initiated Apr 16, 2010)
StatusTerminated
ReasonCML, Inc. was notified by A-M Systems in a letter dated March 31, 2010 that they were recalling a component of their Xenon Kit (ViroMax viral and bacterial filter) because the filters may have a leak at the seam where the two halves of the filter are bonded together.
Root causeRelease of Material/Component prior to receiving test results
ActionThe firm, A-M Systems, sent two "URGENT: Notification of Product Recall/Correction" letters one dated March 31, 2010 and the other dated April 7, 2010 to customers. The letters described the product, problem and action to be taken by customers. Also enclosed was a list of the affected products that was sold to the consignee. The customer was instructed to cease further distribution or use of any remaining products from the listed lots, immediately notify any customers who received products from the lots, provide A-M Systems with the number of items distributed and still remaining in their
Quantity5520
DistributionNationwide distribution: WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063526A-M SYSTEMS VIROMAX VIRAL AND BACTERIAL FILTER
A-M Systems, Inc., Jun 29, 2007