Recall Z-1605-2010 ยท BRY Cabinet, Table And Tray, Anesthesia

Stryker Communications Corp

Universal Flat Panel Yoke, a component of the Stryker Flat Panel Arm. The yoke is intended to hold a flat panel monitor.

DateMay 14, 2010 (initiated Apr 7, 2010)
StatusTerminated
ReasonSome of the pitch shafts were not properly welded which results in the monitor becoming unlevel.
Root causeNonconforming Material/Component
ActionAn Urgent: Medical Device Recall letter dated April 19, 2010 was sent to customers. It was recommended that customers continue use of the product unless the monitor becomes unlevel. The customer was notified that a Stryker Representative would contact their facility to coordinate a service call to inspect the affected units. Questions are directed to the firm via phone at 972-410-7617 or via email at .
Quantity1498
DistributionWorldwide --US and Canada, Australia, China (Hong Kong), Latin America (Mexico, Chile, Ecuador, Costa Rica) .
FDA recordOpen on fda.gov