Recall Z-1782-2010 ยท MKJ Automated External Defibrillators (Non-Wearable)

Defibtech LLC

ReviveR AED semiautomatic external defibrillator used with the DBP-2800 Battery Pack. Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an accessory to the AED and does not have its own intended use. The AED is intended for use on victims of sudden cardiac arrest when the patient is unconscious, unresponsive and not breathing.

DateJun 10, 2010 (initiated May 21, 2010)
StatusTerminated
ReasonThe AED when used with an affected battery pack, may falsely detect an error condition, cancel charge and not provide therapy
Root causeDevice Design
ActionDefibtech, LLC issued an "Urgent Medical Device Safety Information and Correction" notification dated May 20, 2010 by email and hard copy via US Mail. The notification instructed Consignees on proper identification of affected product and steps to take for product replacement. For further information, contact your distributor or Defibtech Technical Support at techsupport@defibtech.com or call 1-877-453-4507 or 1-203-453-4507, 8:30 AM to 5:00 PM Monday - Friday, EST.
DistributionWorldwide Distribution -- United States, Canada, Australia, Jordan, Spain, Israel, UK, Poland, Italy, Netherlands, Malaysia and Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013896SENTRY SEMIAUTOMATIC EXTERNAL DEFIBRILLATOR; EXTERNAL DEFIBRILLATION PADS (ELECTRODES) DDP-100; BATTERY PACK 1200MAH; BA
Defibtech, LLC, Jun 19, 2002