Recall Z-1905-2010 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

Unicel DxH 800 Coulter Cellular Analysis System, Part number(s): 629029

DateJun 25, 2010 (initiated Jul 28, 2009)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter confirmed an issue with the UniCel DxH 800 during in-house testing. It was determined that there is a potential for a false increase in platelet results, due to micro-bubbles. This situation could present a risk of falsely increased platelet results in patient samples as diluent containers are depleted.
Root causeOther
ActionBeginning on July 29, 2009 Customers were contacted by Customer Technical Support (CTS) personnel and provided with the Product Corrective Action (PCA) letter. The PCA letter informed customers of the issue, the impact of the issue and actions/resolutions required by Beckman Coulter Service representative ad the affected consignee. The consignees were informed that their Beckman Coulter Service representative has made temporary changes to the Supplies setup. In addition to the Supplies setup change, consignees were instructed that one of the following workflow options must be implemented to e
Quantity7
DistributionNationwide to: AZ, LA, IL, NY, PA & TN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081930UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Beckman Coulter, Inc., Dec 19, 2008