Recall Z-2179-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Medical Systems Inc

Toshiba America Medical Systems Inc. SSH-880CV Aplio Artida; Diagnostic Ultrasound System

DateMay 10, 2011 (initiated Apr 20, 2010)
StatusTerminated
ReasonThere is a possibility of the system shutting down owing to the power supply unit causing the protection function to activate thus turning off the system prematurely.
Root causeDevice Design
ActionThe recall was initiated on 4/20/10 with Toshiba America Medical Systems (TAMS) forwarding a Medical Device Correction with a Customer Reply Form (via US Postal Service for US Customers) to all customers who purchased the SSH-880CV Aplio Artida Diagnostic Ultrasound System. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to avoid this problem until corrective action is taken, in case this problem occurs; please contact their Toshiba Representative. If emergency use is necessary, use the following procedure
Quantity14
DistributionUSA, including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090158APLIO ARTIDA V2.0 SSH-880CV
Toshiba America Medical Systems, In.C, Feb 6, 2009