Recall Z-1661-2010 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Ge Healthcare, Llc

GE Healthcare, Innova 2100IQ (The Digital Fluoroscopic Imaging System consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)

DateMay 26, 2010 (initiated Mar 24, 2010)
StatusTerminated
ReasonA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios that will cause the next exam to fail and the inability to recall acquired sequences.
Root causeSoftware design
ActionConsignees were sent on 3/24/10, a GE Healthcare " Urgent Medical Device Correction" letter dated March 23, 2010. The letter included the Safety Issue, Affected Product Details, Product Correction and Contact Information.
Quantity92
DistributionWorldwide distribution: USA, Puerto Rico, VENEZUELA, UNITED KINGDOM, TURKEY, THAILAND, TAIWAN, PROVINCE OF CHINA, SWEDEN, SPAIN, RUSSIAN FEDERATION, ROMANIA, POLAND, PHILIPPINES, NORWAY, NETHERLANDS, MOROCCO, REPUBLIC OF KOREA, KAZAKSTAN, JAPAN, ITALY, ISLAMIC REPUBLIC OF IRAN, INDONESIA, INDIA, GER
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050489INNOVA 2100
Ge Medical Systems, LLC, Mar 11, 2005
K052157INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ
Ge Medical Systems, LLC, Aug 25, 2005
K052412INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ
GE Healthcare, Sep 16, 2005
K061163INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT
Ge Medical Systems, LLC, Jun 7, 2006