ClearanceBrief
Recall Z-0314-2011 ยท HAW Neurological Stereotaxic Instrument

FHC, Inc.

Depth Stop Adapter 66-AC-DS(1.8), E6-01-.5 used with the microTargeting and STar Drives Product is the FHC DBS Depth Stop Adapter, a component of the microTargeting Drive and the STar Drive. The Drives are intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other ins

DateNov 3, 2010 (initiated Mar 30, 2010)
StatusTerminated
ReasonFixation thumbscrew on the DBS depth stop adapter may be overtightened and damage the implantable lead.
Root causeDevice Design
ActionFHC issued an "URGENT: MEDICAL DEVICE CORRECTION" letter dated March 29, 2010, to all customers. The letter described the product, problem and actions to be taken by the customers. FHC informed the customers that that they had returned to the original design of the fixation screw in the Depth Stop Adapter. FHC sent replacement Depth Stop Adapter with an original fixation screw. The customers were instructed to discard the recalled Depth Stop Adapter (s) and replace with the enclosed corrected version(s) and sign and fax the receipt notification to 207-666-8309 as proof of their receipt of no
Quantity37
DistributionWorldwide distribution: USA in the states of AZ, CA, FL, ID, NC, MA, MI, MN, NV, RI, UT, VA, and WA, and countries of Columbia and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003776MICROTARGETING DRIVE SYSTEM
FHC, Inc., Feb 23, 2001
K092562MICROTARGETING STAR DRIVE SYSTEM, COMPONENTS
FHC, Inc., Sep 18, 2009