Recall Z-1807-2010 ยท HRS Plate, Fixation, Bone

Acumed LLC

Ulnar Shorting Saw Blade is labeled in part: "Congruent Bone Plates, SIZE Ulnar Shortening Saw Blade - Hub Style S, REF 80-0499-S, Material: Stainless Steel, STERILE, Rx Only, ACUMED, Hillsboro, OR, USA". The Ulnar Saw Blade is an accessory used with a Congruent Bone Plate System covered under K012655, but it is labeled and sold separately by itself. The saw blade is also sold as part of the

DateJun 16, 2010 (initiated Apr 14, 2010)
StatusTerminated
ReasonThe saw blades may not cut aggressively enough for their intended application therefore creating a risk that the saw could generate excessive heat, which may contribute to bone necrosis and/or non-union.
Root causeOther
ActionOn April 14, 2010, Acumed began contacting consignees by telephone. On April 15, 2010, Acumed sent out the URGENT NOTICE: DEVICE RECALL Initial Notification letter and Recalled Product Status Form to their consignees. The letter describes the recall product as Ulnar Shorting Saw Blade-Hub Style S, Part no. 80-0499-S and Lot Codes. The Initial Notification letter instructs consignees to return the product back to Acumed immediately and notify their sub-accounts (customers) if products have been further distributed. The consignees also are asked to fill out and send back the Recalled Produ
Quantity718 units
DistributionWorldwide distribution: USA and to the following countries: Spain, Great Britain, South Africa, Puerto Rico, Canada, Turkey, Ireland, Finland, Belgium, Australia, Germany, and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012655CONGRUENT BONE PLATE SYSTEM
Acu Med, Inc., Nov 7, 2001