Recall Z-1814-2010 ยท DPZ Oximeter, Ear

Ge Healthcare, Llc

DATEX-OHMEDA TRUSAT PULSE OXIMETER AND ACCESSORIES, MODEL 3500

DateJun 22, 2010 (initiated Apr 29, 2010)
StatusTerminated
ReasonGE Healthcare has recently become aware that the TruSat Pulse Oximeter with external power supply does not comply with current safety standards. It is possible that the external power supply could leak electrical current. This current could cause third degree burns or an abnormal heart rhythm that, if untreated, could lead to death.
Root causeComponent design/selection
ActionConsignees were sent on 4/29/10 a GE Healthcare "Urgent Medical Device Correction" letter dated 4/22/10. The letter was addressed to Healthcare Administrator/Risk Manager, Chief of Nursing, Director of Biomedical Engineering, and Home Healthcare Provider. The letter included the Safety Issue, Affected Product Details, Product Correction, Safety Instructions and Contact Information.
Quantity28,442 units
DistributionWorldwide distribution: USA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH,BELGIUM, BOLIVIA, BRAZIL, CANADA, INDIAN OCEAN TERRITORY, CHILE COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONDURAS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040831DATEX-OHMEDA TRUSAT PULSE OXIMETER AND ACCESSORIES, MODEL 3500
Datex-Ohmeda, Inc., Sep 2, 2004