ClearanceBrief
Recall Z-1665-2010 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Biomet, Inc.

Mallory Head Calcar Provisional, 34 mm replacement stem, 11 mm x 220 mm left, Sterile, REF 104773, Biomet Orthopaedics, Warsaw, IN. Implant device intended for use in femoral revision surgery.

DateMay 26, 2010 (initiated Mar 10, 2010)
StatusTerminated
ReasonThe firm has found that some lots of Mallory Head Revision Stems had flute geometry that was out of specification. The flute geometry was too deep.
Root causeProcess change control
ActionBiomet Orthopedics issued an "Urgent Medical Device Recall Notice" dated March 10, 2010. The letter went to distributors and users of the recalled device. The letter instructed the consignees to locate the recalled product and return the product to the firm. For further information, contact Biomet at 1-800-348-9500 or 1-574-372-3983, Monday through Friday, 8:00 AM to 5:00 PM.
Quantity3
DistributionWorldwide Distribution -- United States (OH, IN, FL, KY, NV and CA), Mexico, Chile, China, Japan and The Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031693MALLORY/HEAD MODULAR CALCAR WITH AND WITHOUT HA
Biomet, Inc., Aug 6, 2003