Recall Z-0610-2011 ยท LXH Orthopedic Manual Surgical Instrument

Zimmer Inc.

NexGen Complete Knee Solution Rotating Hinge Knee tibial broach impactor, Catalog # 00-5881-080-00, Zimmer Inc., Warsaw, IN.

DateDec 13, 2010 (initiated Apr 16, 2010)
StatusTerminated
ReasonThe impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery. These complications may also require revision surgery.
Root causeDevice Design
ActionZimmer issued "Urgent Medical Device Safety Alert" letters dated April 16, 2010 to Risk Managers, user facilities, Health Care Professionals and Zimmer Sales force. The users were informed to inspect the device for loosening. The alert also informed users to inspect the device after each use to verify that the device was intact. Customers were instructed to complete and return a User Facility Documentation Form. Zimmer will repair and replace all affected devices. Customers can contact ZImmer, Inc. at 800 613-6131 concerning this recall.
Quantity10,666 both products
DistributionWorldwide Distribution: Throughout US including Puerto Rico, and the countries of Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Denmark, Ecuador, El Salvador, France, Germany, Guatemala, Hong Kong, India, Italy, Japan, Korea, Mexico, Netherlands, Nicaragua, Panama
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.