Recall Z-2482-2010 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Abbott Laboratories, Inc
ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In Vitro Diagnostic use only. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests, utilizing photometry and potentiometric technology, and Immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detect
| Date | Sep 23, 2010 (initiated Oct 21, 2008) |
| Status | Terminated |
| Reason | During manufacturing, two components of the ARCHITECT i1000SR were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover. |
| Root cause | Mixed-up of materials/components |
| Action | ABBOTT issued Technical Service Bulletins to Field Service Representatives who then inspected every unit . |
| Quantity | 404 units distributed worldwide (including US) |
| Distribution | Worldwide distribution:USA including states of: CO, DC, FL, GA, IL, LA, MA, MN, NY, OH, OK, TX, UT, and WV and countries including: Australia, Brazil, Bogota, Canada, China WFOTC, Columbia, Curacao (NETH ANT), Germany, Hong Kong, Japan, Mexico, Singapore, and Taiwan. |
| FDA record | Open on fda.gov |