Recall Z-1527-2010 ยท BSZ Gas-Machine, Anesthesia

Ge Healthcare, Llc

GE, 1009-9002-000 Avance Anesthesia Machine and Monitor

DateMay 11, 2010 (initiated Mar 12, 2010)
StatusTerminated
ReasonUnanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic delivery, and potentially patient monitoring.
Root causeProcess design
ActionConsignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 12, 2010. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager, Director of Biomedical/Clinical Engineering. The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity18 (4 USA, 14 OUS)
DistributionWorldwide distribution: USA (CA, FL, TN), CANADA, DENMARK, FINLAND, FRANCE, ITALY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081844GE DATEX-OHMEDA AVANCE
Datex-Ohmeda, Inc., Aug 29, 2008