Recall Z-3185-2011 ยท MHZ Stimulator, Vestibular Acceleration, Therapeutic

Spacelabs Healthcare, Incorporated

The 91220 mCare 300 Vital Signs monitor has a 10.4 inch resistive touchscreen display. The Spacelabs Medical, mCare 300, Model 91220, Vital Signs Monitor is a compact monitoring system with a resistive touchscreen display. It is indicated for use in adult, pediatric and neonate patient populations wherever there is a need for the monitoring of ECG, respiration, invasive or noninvasive blood press

DateSep 12, 2011 (initiated Apr 14, 2010)
StatusTerminated
ReasonThere have been reports that the Spacelabs Medical Vital Signs Monitor, Model 91220, will not power up with AC power and will no longer charge the batteries.
Root causeComponent design/selection
ActionSpaceLabs Healthcare sent an URGENT - MEDICAL DEVICE CORRECTION letter dated April 15, 2010, to all affected customers. International consignees were notified by e-mail on April 22, 2010, with the customer notification letter. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to weigh the benefits versus the risks when deciding whether or not to continue to use the device until their monitors can be upgraded. Customers were asked to brief their staff about the possibility of the monitor failure to help prevent an adverse
Quantity614 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of UNITED ARAB EMIRATES, ARGENTINA, AUSTRALIA, BELGIUM, BAHRAIN, BRAZIL, CANADA, CHINA, COLOMBIA, GERMANY, ECUADOR, EGYPT, FINLAND, FRANCE, UNITED KINGDOM, GREECE, HONG KONG, JORDAN, REPUBLIC OF KOREA, KUWAIT, MEXICO, MALAYSIA, OMAN, POLAND
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082045MODIFICATION TO MCARE 300 VITAL SIGNS MONITOR, MODEL 91220