ClearanceBrief
Recall Z-1743-2010 ยท KGC Tube, Gastro-Enterostomy

Argon Medical Devices, Inc

ARTERIAL LINE KIT REF/REORDER NO. 498412B DISPOSABLE SINGLE USE, Rx. Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751.

DateJun 2, 2010 (initiated Mar 29, 2010)
StatusTerminated
ReasonThere are 1-2 places on some of the multi-purpose kit blisters/trays in which there are holes and/or the material is so thin that holes are easily created. This would potentially cause the sterility of the products to be compromised.
Root causePackage design/selection
ActionArgon Medical Devices issued an "Urgent Medical Device Recall" notification via certified mail dated March 29, 2010 to distributors and end users. A Response Form was attached. Consignees were informed of the affected product and asked to quarantine and return all unused product to the firm. For further information, contact Argon Medical Devices at 1-903-677-9375.
Quantity1 kit lot #99716572; 60 kits lot #99738188
DistributionDistributed to United States (IL, AL, NY, TX, OK, MI, NC, CA, OH, VA, KY, CO, LA, WA, MO, AND IA) and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K780871ARTERIAL LINE KIT
Argon Medical Corp., Jun 14, 1978