Recall Z-2633-2011 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Tubing Assembly #142, V2, V12 Mixer/Wash a component of the DxC instruments. Tube, Assembly, #142 V2, V12 Mixer/Wash delivers appropriate amount of wash concentrate on the UniCel DxC SYNCHRON Clinical Systems. The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, t

DateJun 23, 2011 (initiated Jul 23, 2009)
StatusTerminated
ReasonThe recall was initiated because the Wash Concentrate Tube Assembly 142 may fail to deliver the appropriate amount of wash concentrate on some SYNCHRON DxC Systems due to an accumulation of crystals. The impact of this issue would include: (1) Quality Control and patient results can potentially be affected. (2) If system has been calibrating properly and Quality Control is recovering within es
Root causePending
ActionBeckman Coulter sent out an "Urgent Product Corrective Action" letter dated July 23, 2009 and mailed the week of July 27, 2009 to all affected customers. The letter provides the cusotmers with explanation of the problem identifed and asks them to take the following actions: " If consignees are not able to resolve these issues through routine troubleshooting they were told to contact BCI Customer Technical Support at (800) 854-3633 in the United States and Canada. " This issue will be corrected with an upcoming software version and tubing modification. " Complete and return the enclosed respo
Quantity2392 units
DistributionWorldwide Distribution - USA (Nationwide) and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042291UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS
Beckman Coulter, Inc., Nov 12, 2004
K060256UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM
Beckman Coulter, Inc., Feb 27, 2006