Recall Z-1437-2010 ยท GWE Stimulator, Photic, Evoked Response

Diopsys, Inc.

Diopsys Inc., Enfant Pediatric VEP Vision Testing System, Input: 120 Vac, 60Hz/2.5A IEC 60601-1 Class I, Type BF.

DateApr 26, 2010 (initiated Apr 2, 2007)
StatusTerminated
ReasonThe computer based system may lose test counts when the time in the device was changed during a changeover from standard time to daylight savings time in March 2007.
Root causeSoftware design
ActionField Service personnel were sent an e-mail notification on 4/2/2007 which included a TZedit file to be downloaded. E-mails instructed reps to install the patch on their customers devices immediately. Field correction was completed 7/31/2007.
Quantity175 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043491ENFANTTM
Diopsys, Inc., Mar 16, 2005