Recall Z-2117-2010 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Dimension Vista LOCI Reaction Vessels, catalog number KS855. For in vitro diagnostic use.

DateAug 2, 2010 (initiated Mar 15, 2010)
StatusTerminated
ReasonVessel pick up errors and vessel jams when the vessel is being loaded. The error halts the instrument and disrupts processing. Vessel flange is thin.
Root causeProcess design
ActionSiemens Healthcare Diagnostics issued an Urgent Field Safety Notice dated March 2010 informing users of the potential problem, to discard any remaining product, and if needed, to request a replacement lot.
Quantity5544 bags
DistributionWorldwide distribution: USA, Canada, Europe, Japan, Korea, and New Zealand.