Recall Z-2212-2011 ยท LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use

Beckman Coulter Inc.

AutoMate TM Sample Processing System, P/N A20745 Basic Assembly - Automate The AutoMate system prepares sample tubes for diagnostic testing and for storage. The AutoMate system contains an automated track that connects to Input/Output, Centrifuge, Serum Level Detection, Decapper, and Output modules. The AutoMate TM 800 model also contains and Aliquot module.

DateMay 19, 2011 (initiated May 7, 2009)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has confirmed an issue with the AutoMate Sample Processing Systems. If during the recovery from Error 81201001, a stuck sample carrier with a tube is moved backwards to the loading position, sample tube/input tray information will become mismatched for samples being loaded onto the AutoMate.
Root causeOther
ActionBeckman Coulter sent an Urgent Product Corrective Action letter dated May 7, 2009, with a Fax Back Response Form (sent by US mail) to all customers who received the AutoMate Sample Processing System. The letter identified the product, the problem and a workaround to correct the problem. Customers were instructed to take the following steps: If the stuck sample carrier is moved backwards during recovery from this error, the sample tube/input tray information will become mismatched. Figure 2 in the PCA letter shows the AutoMate viewed from the rear. Sequence 1 to 3 depicts the error recov
Quantity5 units total
DistributionNationwide Distribution including VA KY and CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.