Recall Z-0107-2011 ยท CEW Radioimmunoassay, Parathyroid Hormone

Beckman Coulter Inc.

Active I-PTH Enzyme-Linked Immunosorbent Assay(ELISA), Part Number: DSL-10-8000 The DSL-10-8000 ACTIVE I-PTH Enzyme-Linked Immunosorbent Assay (ELISA) Kit provides materials for the quantitative measurement of I-PTH in serum or plasma. This assay is intended for in vitro diagnostic use as an aid in the diagnosis of hyperparathyroidism

DateOct 20, 2010 (initiated Mar 4, 2010)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter confirmed that the stated stability of the DSL-10-8000 I-PTH ELISA assay may not be achieved for lot numbers: 891198, 990980 and 991448. The kit lot numbers listed above passed the stability testing performed at 6 months, but failed the testing performed at 12 months. Therefore the affected I-PTH ELISA kit lots may produce falsely elevated patient r
Root causeOther
ActionBeckman Coulter sent an URGENT: PRODUCT CORRECTIVE ACTION letter dated March 17, 2010 with attached Customer Response form to all customers who received Active I-PTH ELISA kits with the affected lot numbers. The letter identified the product, the problem, and the actions to be taken. Customers were instructed to: (1) Discontinue use of all materials provided with I-PTH ELISA kit lot numbers 891198, 990980 and 991448; (2) Review historical external quality control results to ensure the assay performance of these lots was within laboratory specifications. Retesting may be necessary if your l
Quantity70 kits
DistributionNationwide Distribution including the states of CA, CT, IA, IL, MI, NY, SC, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952199ACTIVE INTACT PTH ELISA
Diagnostic Systems Laboratories, Inc., Jul 17, 1995