Recall Z-2621-2010 ยท CGR Radioimmunoassay, Cortisol

Beckman Coulter Inc.

Access Immunoassay Systems CORTISOL, Part Number: 33600

DateSep 29, 2010 (initiated Feb 17, 2010)
StatusTerminated
ReasonThe recall was initiated by Beckman Coulter after the firm confirmed customer reports that the affected lot of Access Cortisol reagent (REF 33600) identified above may produce suppressed patient results. Beckman Coulter also confirmed customer reports that the two lots of Access Progesterone reagent (REF 33550) identified above may produce suppressed patient results and an increased incidence of I
Root causeOther
ActionTwo (2) Product Corrective Action (PCA) letters with attached Customer Response form were sent on the week of February 15, 2010 to the affected customers. The letters provide the customers with an explanation of the problem identified and instruct the customers to discontinue use of Access Progesterone and Access Cortisol Reagents, lot numbers referenced. For product replacements consignees were instructed : (1) In the United States, please contact Customer Service at 1-800-526-3821; (2) In Canada, please contact Customer Service at 1-800-463-7828; (3) Outside of the United States and Cana
Quantity7282 units
DistributionUnited States & Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K954733ACCESS CORTISOL ASSAY (MODIFICATION)
Bio-Rad Laboratories, Inc., Dec 5, 1995