Recall Z-0456-2011 ยท CGL Electrode, Ion Based, Enzymatic, Creatinine

Abbott Point of Care Inc.

i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.

DateNov 24, 2010 (initiated Mar 17, 2010)
StatusTerminated
ReasonThe i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.
Root causeOther
ActionAbbott Point of Care issued Urgent Product Recall Notification letters dated March 2010 to all direct customers identifying the affected product and the actions to be taken. Customers were instructed to determine if they had cartirdges that had been stored above 25 degrees Celsius, and if so, contact Abbott for replacement or credit. Abbott Technical Support can be contacted at 800-366-8020 option 1 about this recall.
Quantity34525 cartridges
DistributionWorldwide Distribution: Throughout USA and Canada.