Recall Z-1758-2010 ยท JQW Station, Pipetting And Diluting, For Clinical Use

Beckman Coulter Inc.

COULTER PrepPlus, Part Number: 286600

DateJun 4, 2010 (initiated Jan 21, 2010)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter confirmed reports of potential dilution of reagent vials loaded onto the Preppies or PrepPlus 2 systems. Affected systems have a syringe pump with part number A46748 (see illustrations below) which cause the probe to prematurely dispense Isoflow prior to arriving at the wash station. This condition only occurs with worklist panels which include both
Root causeOther
ActionA Product Corrective Action (PCA) letter with attached PCA Response Form was sent on the week of January 25, 2010 to all customers who purchased the pipetting and diluting system: COULTER PrepPlus, or COULTER PrepPlus 2. The letter provides the customers with an explanation of the problem identified and an instruction on proper placing antibody reagent when using a worklist panel which involves both blood specimens and antibody reagents. Customers were also instructed to complete and return the PCA Response Form in addition to sharing the information contained in the PCA with their laboratory
Quantity193 units in total
DistributionDomestically and Canada.