Recall Z-1596-2010 ยท LOL Hepatitis A Test (Antibody And Igm Antibody)

Abbott Diagnostics International, Ltd.

ARCHITECT HAVAB-M Reagent A chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgM antibody to hepatitis A virus (IgM anti-HAV) in human adult and pediatric serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum.

DateMay 19, 2010 (initiated Dec 30, 2009)
StatusTerminated
ReasonThere is a potential to generate a false grayzone or reactive HAVAB-M result when the ARCHITECT AUSAB assay precedes the ARCHITECT HAVAB-M assay.
Root causePending
ActionDirect accounts (hospitals, private laboratories, and other institutions) were sent a Product Correction letter, dated 12/30/2009. The letter identified the affected product, explained the issue and patient impact, and listed several necessary actions. Customers were instructed to follow one of 4 different alternatives to protect the integrity of test results. The letter is to be retained for laboratory records and a copy should be sent to other laboratories that kits were forwarded to. Questions regarding this information should be directed to Customer Service at 1-877-4ABBOTT.
Quantity1,326
DistributionNationwide Distribution -- Including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063329ABBOTT ARCHITECT HAVAB-M
Abbott Laboratories, Apr 16, 2007