Recall Z-2771-2011 ยท LIG Radioassay, Intrinsic Factor Blocking Antibody

Beckman Coulter Inc.

UniCel DxC 880i, 860i,680i and 660i Synchron Access Clinical Systems The Unicel DxC Integrated Workstation takes samples loaded from a single point of entry through a UCTA (UniCel Closed Tube Aliquotter). The UCTA functions as the sample processing manager by aliquotting and routing samples to the DxC and the Dxl analyzers according to the programming requirements

DateJul 11, 2011 (initiated May 7, 2009)
StatusTerminated
ReasonThe recall was initiated after Beckman Coulter confirmed reports of inaccurate Access Intrinsic Factor Ab test results displayed on, or printed from, the UniCel Dxl portion of an integrated workstation listed above. The problem is limited to numerical Intrinsic Factor Ab results produced on integrated systems running UniCel Dxl system software version 4.0, 4.1, or 4.2. The DxC portion of your inte
Root causeSoftware design
ActionBeckman Coulter, Inc. sent out an "Urgent: Product Corrective Action" letter dated May 19, 2009 to all affected customers. The letter explained the reason for recall and provided recommendations including: -System software currently under development will correct this problem, and will be made available soon. Until then, please refer to UniCel DxC reports, user interface information, and LIS transmissions, which contain accurate Access Intrinsic Factor Ab test results. -Refer to the UniCel DxC system documentation for information on displaying and printing results from the DxC por
Quantity5 units
DistributionWorldwide Distribution: (USA) Nationwide including KY; and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060256UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM
Beckman Coulter, Inc., Feb 27, 2006