Recall Z-2267-2010 ยท FNL Bed, Ac-Powered Adjustable Hospital

Hill-Rom, Inc.

Hill-Rom Care Assist Bed, Model P1170.

DateAug 24, 2010 (initiated Apr 13, 2009)
StatusTerminated
ReasonPinch hazard. Potential for shearing action between the side rail arm and the bracket used to secure the side rail to the bed.
Root causeDevice Design
ActionHill-Rom sent a customer notification letter dated 4/9/2009. A second notice was sent to all customers on 7/9/2009. The letter included instructions and a Customer Response Form, along with the required parts. Customers were instructed to do the correction on their affected beds and return the completed customer response form within 45 days.
Quantity39,971 both products
DistributionWorldwide distribution: USA, Afghanistan, Algeria, American Samoa, Argentina, Australia, Bermuda, Brazil, Canada, Chili, China, Colombia, Costa Rica, Cuba, Ecuador, Egypt, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Kyrgyzstan, Lebanon, Libya, Malaysia, Mexico, Oman, Panama,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.