Recall Z-2047-2011 ยท CFT Radioimmunoassay, Prolactin (Lactogen)

Beckman Coulter Inc.

Active Prolactin ELISA, 96 Wells, DSL-10-4500 The DSL-10-4500 Active Prolactin Enzyme-Linked Immunosorbent Assay (ELISA) Kit provides materials for the quantitative measurement of prolactin in human serum. This assay is intended for in vitro diagnostic use in the diagnosis and treatment of disorders of the anterior pituitary gland or of the hypothalamus portion of the brain.

DateApr 20, 2011 (initiated Aug 3, 2009)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has confirmed that the Active Prolactin ElLISA kit, lot number 890291 over estimates the expected dose values for patient serum samples due to a shift in the assigned calibrator values. Beckman Coulter recommends that patient results generated with the affected lot be evaluated in the context of other diagnostic tests and the clinical presentatio
Root causeEquipment maintenance
ActionThe firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION" Letter dated August 03, 2009 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: Action/Resolution: Discontinue use and discard remaining product with lot number 890291 of Active Prolactin reagents. Contact Beckman Coulter at 1-800-231-7970 for replacement product. Newer lots (lot number 991210 or higher) of the product have been realibrated to restore the performance as described in the Instructions for Use (IFU). For method correlation
Quantity30 Kits
DistributionNationwide distribution: USA including states of: FL, GA and NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960561DSL ACTIVE PROLACTIN ELISA MODEL 10-4500
Diagnostic Systems Laboratories, Inc., Apr 26, 1996