Recall Z-2241-2010 ยท NRU Chair, Dental, Without Operative Unit
Dental Equipment LLC
The magnetic headrest is designed to fit the following dental chairs (THE HEADREST IS THE RECALLED ITEM, NOT THE CHAIRS): The dental chairs are sold under the brand names Marus and DCI Equipment. The affected models are: Marus MaxStar-Agency Model # DC1490; Marus MaxStar-Agency Model # DC1690; Marus NuStar-Agency Model # DC1700; Marus NuStar-Agency Model # DC1702; Marus ProStar-Agency Model
| Date | Aug 20, 2010 (initiated Mar 23, 2010) |
| Status | Terminated |
| Reason | The magnetic headrest on certain dental chairs may affect the function/programming of some implantable pacemakers or defibrillators if the implanted device is programmed to respond to a magnet. |
| Root cause | Other |
| Action | DCI Equipment sent an "Urgent Field Safety Alert" letter dated March 15, 2010 to consignees advising them of the potential problem. The letter instructed that patients with implantable pacemakers or defibrillators should avoid dental chairs with the magnetic headrest. A warning label was sent with the letter for insertion into the product Use and Care manual for operator warning. Consignees were asked to complete a Field Safety Alert Acknowledgement/Return Form and fax to the firm. Consignees were requested to notify customers.Consignees may contact the firm at 503 537-3602. |
| Quantity | 108 |
| Distribution | Worldwide distribution in the USA and Canada. |
| FDA record | Open on fda.gov |