ClearanceBrief
Recall Z-1452-2010 ยท MDS Sensor, Glucose, Invasive

Abbott Diabetes Care, Inc.

Freestyle Navigator Continuous Glucose Monitor System, ART 16467 Rev A, System, Business 1, MMOL. Abbott Item Number 70966-02, distributed by Abbott Diabetes Care Inc, Alameda, CA. Medical device indicated for continually recording interstitial fluid glucose levels in people ages 18+ with diabetes mellitus for improving diabetes management. The device is indicated to be used in hospital and ho

DateApr 30, 2010 (initiated Apr 14, 2009)
StatusTerminated
ReasonPotential failure or inaccurate readings--the device's plastic housing near the battery compartment may crack, and allow moisture to enter.
Root causeProcess control
ActionNotification letters were issued to FreeStyle Navigator customers via certified mail and notification was posted on the Abbott Diabetes Care's Website. Customers were notified of the potential issue and actions to take: 1) if moisture is observed in the FreeStyle Navigator transmitter now or at any time, discontinue use of the transmitter and call Abbott Diabetes Care Customer Care at 1-800-418-6427 and 2) if a loss of connection is encountered or if your results are not reliable or are inconsistent with how you feel, consult your User's Guide for troubleshooting guidelines. If the issue i
Quantity5,449 of total units
DistributionWorldwide Distribution -- United States (OH, FL, NJ, CA, MS, MN, MA, and NY), France, Australia, the Netherlands, Sweden, the UK, Israel, Ireland and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.