Recall Z-1429-2010 ยท IWB System, Radiation Therapy, Radionuclide

Elekta, Inc.

Leksell Gamma Knife C

DateApr 22, 2010 (initiated Aug 1, 2006)
StatusTerminated
ReasonDue to a material change in a critical component of the linear actuators the quality of said part was found to be inadequate.
Root causeOther
ActionPhase 1 - Technical Notes Leksell Gamma Knife C version 1.2 (dated August 2006) "Potential Problem with Error Handling" was sent to all users. Phase 2 - Mandatory Field Action 0000748, "Helmet Changer Actuator, MTB and MCU SW Upgrade Kit was released December 3, 2007. This Field Change Order includes two mod kits and a software upgrade to provide a permanent solution to the issues. Phase 3 - Technical Note 200 068, "Safety issue with helmet host actuator", released on May 15, 2008. This Technical Note includes a warning to customers of a potential quality problem with the helmet hoist a
Quantity51 units
DistributionAL, CA, CO, CT, FL, GA, IL, MD, MA, MI, MN, MO, NE, NY, NC, OH, OK, OR, PA, TN, TX, VA, WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984328LEKSELL GAMMA KNIFE TARGET SYSTEM, MODEL 24001
Elekta Instrument AB, May 21, 1999