Recall Z-1428-2010 ยท HAW Neurological Stereotaxic Instrument

Elekta, Inc.

Leksell SurgiPlan

DateApr 22, 2010 (initiated May 23, 2005)
StatusTerminated
ReasonImages from a different patient can be imported without a warning for the user.
Root causeOther
ActionPhase I - A technical note for LGP and SPS describing the risk for importing wrong image files and including recommendations on how to reduce this risk has been written and distributed to users of both Leksell GammaPlan and Leksell SurgiPlan on May 23, 2005 via FCO DOC 05-0001. Phase 2 - A maintenance release was created of the Scanner Interface (SIF) to correct the problem. This was distributed to Leksell SurgiPlan users on November 30, 2005 via FCO0000153-1. Additional questions are directed to the firm to Product Manager lrobert.aslund AT elekta.com, or After Sales gustaf.piehl AT elek
Quantity57 units
DistributionNationwide, Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943468LEKSELL SURGIPLAN
Elekta Instruments, Inc., Jan 20, 1995