Recall Z-1323-2010 ยท DTN Reservoir, Blood, Cardiopulmonary Bypass

Circulatory Technology Inc

The Bigger Better-Bladder, Catalog Number BBB38 (with 3/8 inch ID tubing), Sterile, Single use only --- Manufactured for: Circulatory Technology Inc., 21 Singworth Street, Oyster Bay, 11771 (516) 624-2424 --- 510 (k) 981284.

DateApr 13, 2010 (initiated Jun 3, 2009)
StatusTerminated
ReasonDuring an FDA inspection, it was discovered that the firm had conducted a silent recall of its Bigger Better Bladder device after receiving a complaint from a hospital that the device collapsed during a procedure thereby obstructing flow.
Root causeProcess control
ActionThe firm, Circulatory Technology, Inc., sent an email to Bigger Better-Bladder end user accounts on 6/03/09 and again on 6/13/09. The firm followed up with phone calls to the user customers. The emails and calls discussed the product, problem and the actions to be taken by the customers. The customers are ask to examine their stock of Bigger Better-Bladder for the affected units and then visually examine each unit to assure that adhesive has been applied and when the tube is bent away from cap, no separation is visible along the joint between the two according to the picture provided. The c
Quantity144 units
DistributionArizona, Arkansas, California, Florida, Indiana, Iowa, Maine, Michigan, Nebraska, New York, North Carolina, Pennsylvania, Ohio, Texas, Utah, Washington DC, and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981284BETTER-BLADDER (BB)
Circulatory Technology, Inc., Aug 12, 1998