Recall Z-0501-2011 ยท HWN Instrument, Compression
Zimmer Inc.
Zimmer Reconstruction System Offset Clamp Tips 4.5 cm Offset, Item # 00-1179-028-03, Zimmer Inc., Warsaw, IN. Intended to temporarily reduce pelvic fractures for fixation during reconstructive surgery.
| Date | Dec 1, 2010 (initiated Feb 16, 2010) |
| Status | Terminated |
| Reason | The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, increasing the potential for fracture. |
| Root cause | Nonconforming Material/Component |
| Action | Zimmer sent an Urgent Device Recall letter dated February 16, 2010, to all affected customers. The letter identified the product, the problem, and the action the customers should take. Customers were instructed to monitor all potentially affected patients through regular office consultations. Ensure Zimmer Reconstruction System Tenaculum System Instrument Users at their facility were informed of the urgent device recall. Locate all instruments and quarantine them immediately. Count all affected product in their territory and record the data on the Inventory Return Certification Form include |
| Quantity | 217 all products |
| Distribution | Worldwide Distribution - USA including AL, FL, HI, IA, IL, KS KY, LA, MI, MO, MS, NC, ND, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, AND WI and the countries of Australia, Canada, Central or South America, China, Germany, India, Italy, Saudi Arabia, and UK, |
| FDA record | Open on fda.gov |