ClearanceBrief
Recall Z-0497-2011 ยท HWN Instrument, Compression

Zimmer Inc.

Zimmer Reconstruction System Tenaculum Standard 9 inch, Item # 00-1179-025-00, Zimmer Inc., Warsaw, IN. Intended to temporarily reduce pelvic fractures for fixation during reconstructive surgery.

DateDec 1, 2010 (initiated Feb 16, 2010)
StatusTerminated
ReasonThe firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, increasing the potential for fracture.
Root causeNonconforming Material/Component
ActionZimmer sent an Urgent Device Recall letter dated February 16, 2010, to all affected customers. The letter identified the product, the problem, and the action the customers should take. Customers were instructed to monitor all potentially affected patients through regular office consultations. Ensure Zimmer Reconstruction System Tenaculum System Instrument Users at their facility were informed of the urgent device recall. Locate all instruments and quarantine them immediately. Count all affected product in their territory and record the data on the Inventory Return Certification Form include
DistributionWorldwide Distribution - USA including AL, FL, HI, IA, IL, KS KY, LA, MI, MO, MS, NC, ND, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, AND WI and the countries of Australia, Canada, Central or South America, China, Germany, India, Italy, Saudi Arabia, and UK,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.