Recall Z-1423-2010 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo Imaging XS model number 10496279 The intended use: image processing radiological

DateApr 20, 2010 (initiated Feb 16, 2010)
StatusTerminated
Reasonimages can be overwritten
Root causeSoftware design
ActionA "Customer (Safety) Advisory notice" dated February 16, 2010, was sent all affected customers via Update Instructions IM019/10/S. This notice describes the product, problem and the action to be taken by the customers. The customers should immediately instruct their employees of the advisor notice and maintain awareness. If the customer sold the device/equipment and is no longer in possession of it, they should forward the notice and inform Siemens of the new owner. A software update to correct this issue will be distributed and installed by a Siemens Service Engineer on all affected system
Quantity6 units
DistributionNationwide distribution: CA, NC, NY, PA, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082430SYNGO IMAGING XS
Siemens AG Medical Solutions, Nov 14, 2008