Recall Z-1432-2010 ยท JAI Couch, Radiation Therapy, Powered

Elekta, Inc.

Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.

DateApr 23, 2010 (initiated Jan 11, 2007)
StatusTerminated
ReasonIn some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
Root causeOther
ActionMandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.
Quantity185 units
DistributionWorldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serb
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983678ELEKTA ONCOLOGY SYSTEMS PRECISE TREATMENT TABLE
Elekta Instruments, Inc., Jan 15, 1999